A-PULSE CASPal
 
A-PULSE CASPal® Specification
Specifications Parameters
General
Model T7200 series
Display §          51.3 x 75.5 mm (5 lines) monochrome LCD
Battery life Approximately 500 measurements with 4 new standard alkaline “AA” batteries and inflating the cuff to 160mmHg at a room temperature of 22°C
Control buttons (soft key) §          ON/OFF §          BP §          CASP §          MEMORY
Display
Parameters §          Systolic pressure in mmHg §          Diastolic pressure in mmHg §          Pulse rate in BPM (Beats Per Minute) §          Central aortic systolic pressure in mmHg
Pulse Waveform Real-time radial arterial pulse waveform
Plot area range §          Auto-scalable Y-axis, digital value of 0 to 4395 §          X-axis, 10 seconds
Blood Pressure & CASP Measurement
Calibration method Oscillometric, performed before use
Measurement site §          Arm (left or right) for blood pressure measurement §          Radial artery at the wrist (left or right) for wave acquisition
Cuff inflation method Automatic, by electric pump
Cuff deflation method Electronic control valve (modulated)
Rapid pressure release method Electronic control valve (open)
Blood pressure measurement time Less than 45 seconds (typically)
Blood pressure range 40 to 280 mmHg
Measurement accuracy Meets ANSI/AAMI SP10:1992 Standard
Pulse waveform acquisition device Radial pulse wave acquisition device
Pulse waveform acquisition method Modified applanation tonometry
Pulse rate range 30 to 180 beats/minute
Pulse rate accuracy ±3%
Pulse waveform acquisition time Approximate 10 seconds
Memory capacity 200 sets of  blood pressure measurements and CASP readings
Auto power off time 5 minutes of non-use
Dimensions and Weight
Main unit dimensions W117 x D131 x H86 mm (W4,6 x D5,2 x H3,4 inch)
Main unit weight Approximately 330g (12 oz.)
Environmental
Operating conditions ·      10°C to 40°C ·      15% to 90% relative humidity (non-condensing) ·      80 to 106 kPa atmospheric pressure
Storage conditions ·      -20°C to 50°C ·      10% to 95% relative humidity (non-condensing)
Others
Cuffs ·      Regular adult cuff,      for arm circumferences 21 to 30cm (8.3 to 11.8 inches), (included with system)      PART NUMBER: P1500B ·      Large adult cuff, for arm circumferences 29 to 42 cm (11.4 to 16.5 inches) (optional)
PART NUMBER: P1501B
Patient safety Automatic cuff deflation if any of the following conditions is encountered: ·      Cuff pressure exceeds 290 mmHg ·      Measurement time exceeds 180 seconds ·      Safety timer detects microprocessor failure
Regulatory
Electrical safety ·      IEC60601-1:3.Ed.  
EMC ·      IEC60601-1-2:2007   ·      FCC Part 18  
Safety ·      EN 1060-1:2002   ·      EN 1060-3: 1997  
Class ·      Class IIa, in accordance with EC directive 93/42/EEC (Medical Device Directive) Annex IX rule 10, Patient applied part type BF  
Performance ·      ANSI/AAMI SP10:1992  
Environmental ·      MIL-STD810F  
 

Copyright © 2010 HealthSTATS International. All Rights Reserved. Disclaimer & Copyright