A-PULSE CASPro
 
A-PULSE CASPro® Specifications

 

Specifications

Parameters

General

Model

T7100 series

Display

§          5.7 inch (320 x 240 pixels) color TFT LCD

Power source

§          6VDC AC-DC adapter, AC100V - 240V 50/60Hz, or

§          4 x 1.5V (AA size) alkaline batteries, non-rechargeable type

Battery life

Approximately 50 measurements with 4 new standard alkaline “AA” batteries and inflating the cuff to 160mmHg at a room temperature of 22°C

Control buttons (soft key)

§          BP

§          AUTO CASP

§          MEMORY

§          SETUP

Display

Parameters

§          Systolic pressure in mmHg

§          Diastolic pressure in mmHg

§          Central aortic systolic pressure in mmHg

§          Mean arterial pressure in mmHg

§          Pulse rate in BPM (Beats Per Minute)

§          Radial augmentation index in %

§          Radial augmentation pressure in mmHg

§          Peak relative time in mSec (millisecond)

Pulse Waveform

Real-time radial arterial pulse waveform

Plot area range

§          Auto-scalable Y-axis, digital value of 0 to 4395

§          X-axis, 10 seconds

Blood Pressure & CASP Measurement and Pulse Waveform Acquisition

Calibration method

Oscillometric, performed before use

Measurement site

§          Arm (left or right) for blood pressure measurement

§          Radial artery at the wrist (left or right) for wave acquisition

Cuff inflation method

Automatic, by electric pump

Cuff deflation method

Electronic control valve (modulated)

Rapid pressure release method

Electronic control valve (open)

Blood pressure measurement time

Less than 45 seconds (typically)

Blood pressure range

40 to 280 mmHg

Measurement accuracy

Meets ANSI/AAMI SP10:1992 Standard

Pulse waveform acquisition device

Radial pulse wave acquisition device

Pulse waveform acquisition method

Modified applanation tonometry

Pulse rate range

30 to 180 beats/minute

Pulse rate accuracy

±3%

Pulse waveform acquisition time

10 seconds per block of waveform

Memory capacity

200 sets of  blood pressure measurements and CASP readings

Auto power off time

Min. 5 minutes, Max. 15 minutes of non-use  (to be defined by user)

Dimensions and Weight

Main unit dimensions

W160 x D161X H95 mm (W6,3 x D6,3 x H3,7 inch)

Main unit weight

Approximately 0.8 kg with battery (main unit only)

Environmental

Operating conditions

·      10°C to 40°C

·      15% to 90% relative humidity (non-condensing)

·      80 to 106 kPa atmospheric pressure

Storage conditions

·      -20°C to 50°C

·      10% to 95% relative humidity (non-condensing)

Others

Cuffs

·      Regular adult cuff,

     for arm circumferences 21 to 30cm (8.3 to 11.8 inches), (included with system)

     PART NUMBER: P1500B

·      Large adult cuff,

for arm circumferences 29 to 42 cm (11.4 to 16.5 inches) (optional)
PART NUMBER: P1501B

Patient safety

Automatic cuff deflation if any of the following conditions is encountered:

·      Cuff pressure exceeds 290 mmHg

·      Measurement time exceeds 180 seconds

·      Safety timer detects microprocessor failure

Regulatory

Electrical safety

·      IEC60601-1:3.Ed.

EMC

·      IEC60601-1-2:2007  

·      FCC Part 18  

Safety

·      EN 1060-1: 2002

·      EN 1060-3: 1997  

Class

·      Class IIa, in accordance with EC directive 93/42/EEC (Medical Device Directive) Annex IX rule 10, Patient applied part type BF  

Performance

·      ANSI/AAMI SP10:1992  

Environmental

·      MIL-STD810F  

 

 

Copyright © 2010 HealthSTATS International. All Rights Reserved. Disclaimer & Copyright