• Simplified Chinese
   
 
BPro
 
BPro® Specifications Print

Specifications

Parameters

General

Model

T6200 series

Display

LCD, 20 mm X 20 mm 2 lines

Displayed parameters

     - ABPM mode

Systolic pressure in mmHg
Diastolic pressure in mmHg
Pulse rate in beats/minutes

     - PWA mode

Pulsation bar

Memory (For ABPM)

96 blocks of waveform readings

Reading cycle time (For ABPM)

8 to 10 seconds

Measurement range

    - Blood pressure

40 to 280 mmHg

    - Pulse rate

30 to 180 beats/minute

Power source

3V CR 2032 coin battery

Battery life

1 – 6 months depending on usage

Buttons

Five buttons

1. Mode
2. Light
3. Forward
4. Reverse
5. BP

Operating Principles

Measurement method

Modified arteries applanation tonometry

Measurement site

Radial artery at the wrist (Left or Right)

Calibration

Calibration method

Oscillometric, performs prior to use

Size and Weight

Size – Watch head

44mm W X 51mm L X 19mm H

Size – Sensor housing

32mm W X 38mm L X 16mm H

Weight

20 grams

Environmental

Operating conditions

10 to 40ºC, 15 to 90% relative humidity (non-condensing).

80 to 106kPa atmospheric pressure

Regulatory compliance

FDA, CE, SFDA, TGA, HSA

Accessories

BP Calibrators, blood pressure cuff

Regular adult, for arm circumferences 21 to 30 cm. Included with system.

Optional: Large adult, for arm circumferences from 29 to 42 cm.

 

Copyright © 2010 HealthSTATS International. All Rights Reserved. Disclaimer & Copyright